Home baths HC/BW Series Constant-temperature water baths

HC/BW Series Constant-temperature water baths


HC/BW Series Constant-temperature water baths



  • Precise control of temperature can be realized through high-precision PID temperature controller and circulating water agitation.
  • Touch-key is used to set temperature, making operation simple and convenient.
  • LED is used for direct digital display of actual temperature and measured temperature.
  • With bigger opening size, it is convenient for users to make direct thermostatic test of samples, it is possible to install outside circulating pump.
Features:

Standard Configuration

Produce name Water bath Precise water bath
Model HC06 HC010 HC020 BW06 BW010 BW020
Performance
Temperature range R.T +10℃~ 95℃
Temperature constancy ±0.05℃ ±0.01℃
Temperature uniformity Level: ±0.05℃ Largest: ±0.1℃ Level: ±0.02℃   Largest: ±0.04℃
Components
Interior material Stainless steel SUS304
Exterior material Electrostatic spraying of epoxy powder on cold-rolled steel sheet
Heat insulating material Glass wool
Agitation device Gear coupling device
Heater Stainless steel heater pipe
Sensor PT100 Ω
Controller
Temperature control system PlD control by microprocessor
Temperature setting method Digital setting by key
Temperature display method Digital display by LED
Temperature measurement  resolution 0.1℃ 0.001℃
Operation function Fixed temperature operation
Heater Circuit Control Triac zero-cross type
Safety device
Overcurrent breaker, Electric Leakage Breaker
Internal dimensions (mm) D 200 300 400 200 300 400
W 200 250 300 200 250 300
H 150 150 150 150 150 150
Top opening(mm) D 200 300 400 200 300 400
W 200 250 300 200 250 300
Internal capacity Approx.6L Approx.11L Approx.18L Approx.6L Approx.11L Approx.18L
Power Supply AV220V/50HZ (single phase)
Accessories
Stainless steel lid,1pcs/Product description 1 set
Optional Accessories
Programs controller/Pump of output loop etc.


  • The performance shown here is the value under the condition of the ambient temperature of 23∞C±2∞C, humidity of 65%RH±20%. And there is not burden.
  • The projection is not included for Internal dimensions and External dimensions.
 
 
 